Lion Precision offers the ECL150 and ECL150e Multi-Channel Eddy-Current Displacement Sensors. Up to eight channels of high-performance displacement sensing are integrated in a compact package.
Key Features
The main features of the ECL150 & ECL150e Multi-Channel Eddy-Current Displacement Sensors are:
- Linearity: ±0.2%
- Resolution (Typical, Dependent on Probe and Range)
- ECL150 Non-ferrous: 0.006% RMS
- ECL150 Ferrous: 0.008% RMS
- ECL150e: 0.3 µm RMS or higher
- Selectable Bandwidth (kHz): 0.25, 1, 10, 15
- Low power/heat
- Small volume package
- ±5 V DC outputs
- 1-8 Channels
- Range indicating LEDs
- CE Compliant
- Available without enclosure
- Probes can be ordered as vacuum compatible.
Specifications
The specifications of the ECL150 & ECL150e Multi-Channel Eddy-Current Displacement Sensors are as follows:
|
|
|
Resolution:
RMS, Typical
Dependent on probe and range.
See Range Table for resolution specifics.
|
ECL150 |
0.006% RMS (Non-ferrous);
0.008% RMS (Ferrous) |
ECL150e |
Greater of 0.3 µm or
0.006% RMS (Non-ferrous);
0.008% RMS (Ferrous) |
Linearity |
±0.2% of Range |
Error Band |
±0.4% of Range |
Maximum Bandwidth |
15 kHz |
Driver Temperature Drift
(15°C-50°C)
Typical. See Probe/Range chart for details.
|
non-ferrous |
±0.01% F.S./°C |
ferrous |
±0.01% F.S./°C |
Probe Temperature Drift
(15°C-65°C)
Typical. See Probe/Range chart for details.
|
non-ferrous |
±0.01% F.S./°C |
ferrous |
±0.01% F.S./°C |
Analog Output |
±5 VDC, 0 Ù, 15 mA |
Input Power |
12 to 24 VDC,
1-2 Channels – 2.5 W
3-4 Channels – 3.5 W
5-6 Channels – 4.5 W
7-8 Channels – 5.5 W |
Driver Operating Environment |
4°C-50°C
IP64 |
Probe Operating Temperatures
Standard Probes come with PUR (Polyurethane) jacketed cables. High Temperature Probes have FEP jackets.
Other temperature ranges may be possible. Contact us for more information.
Standard Probes: IP67
High Temperature Probes: IP63
|
U3 Probe |
Standard Probe (PUR Cable):
-25°C to +125°C
(Available with optional FEP cable)
High Temperature Version Not Available
|
U5-18 Probes |
Standard Probes:
-25°C to +125°C
High Temperature Probes:
-25°C to +200°C
|
U25-50 Probes |
Standard Probes:
-25°C to +125°C
High Temperature Probes:
-25°C to +175°C
|